Audionvac VMS 363 Validatable

Validatable single chamber with flat plate

The validatable VMS 363 is the largest single chamber vacuum packaging machine with flat plate of the Audionvac range. With the 5 different seal bar configuration and a total sealing length of 1090 mm it runs smoothly.

This strong validatable floor model vacuum packaging machine is available as an industrial model full option vacuum packaging machine. Made from heavy industrial materials which makes it a robust and durable packaging machine.

“Audionvac: professional range of validatable vacuum packaging machines”

  • Complies to ISO 11607-2 validation requirements for sealing
  • Critical process parameters to be controlled and monitored:
    • Vacuum / gas pressure
    • Seal temperature
    • Seal time
    • Seal pressure
  • Equipped with accurate Audion Temperature Controller (ATC)
  • Equipped with user friendly touchscreen interface (ADC)
  • Automatic stop of cycle if one of the critical parameters has exceeded alarm limit
  • Critical process parameters can be calibrated
  • Seal data can be logged and exported to PC via USB stick
  • Access level controllable by login passwords for different user levels
  • Bi-active 8mm seal bars
  • Equipped with accurate Audion Temperature Controller (ATC)
  • Equipped with user friendly touchscreen interface (ADC)
  • Access level controllable by login passwords for different user levels
  • Bi-active seal bars
  • Seal width 8 mm
  • Busch vacuum pump
  • Maintenance friendly and easy to clean construction
  • High quality materials
  • Standard with up to 50 recipe templates
  • Aluminium vacuum chamber
  • Gas flush, modified atmosphere packaging (MAP)
  • Multicycle
  • Unique user ID and password
  • Barcode scanner
  • Label printer
  • Calibration
  • Validation
  • External pump
  • External exhaust
  • Parameter verification
    – Seal temperature verification set (ATM)
    – Seal pressure verification set (ASPM)
    – Vacuum pressure verification set (AVLM)
  • Calibration
    – Audion factory calibration
  • IQ/OQ Check
    – 1 x IQ check for 1 machine
    > Operation training (max. 4 persons)
    > Maintenance training (max. 4 persons)
    – OQ check for 1 type of bag
    > Peel test for defining seal temperature
    > Peel test with defined seal temperature
    > Dye penetration test
    > Seal check
    – Extra OQ check for each extra bag
  • Seal integrity test
    – Seal integrity test service
    > Peel test
    > Dye penetration test
    > Seal check
    – Peel tester
    – Seal check sheet
    – Seal check sheet in roll
    – Blue ink for dye penetration test
  • Pharmaceutical
  • Medical
  • High-Tech
  • Electronics
  • Cleanrooms

Specifications

ProductAudionvac VMS 363 Validatable
Seal length
1x 1090 + 1x 610 mm (S/L)
2x 1090 mm (L/L)
2x 660 mm (S/S)
1x 1090 + 2x 610 mm (S/L/S)
Seal width8 mm
Position of the sealing bar
Effective chamber size (LxWxH)610 x 1050 x 280 mm (S/L)
1090 x 590 x 280 mm (L/L)
660 x 1000 x 280 mm (S/S)
610 x 1000 x 280 mm (S/L/S)
Type of machine
Validatable/calibratable/medical, Vacuum chamber, IQ/OQ
ModelFloor
CoveringStainless Steel
Pump capacity160 m3 /h Busch pump
Pedal operatedNo
Type of filmPA/PE, Vacuum Bags, Aluminum Laminated Bags
Thickness of film2 x 150 micron
Type of packageBag
Branche/ApplicationIndustrial Parts, Medical & Laboratory, Pharmaceutical Industry
Type of productSolid
Machine cycle3 – 5 cycles per minute
Power4,0 – 4,2 kW
Pneumatic supply0.3 MPa
Voltage400V-3-50Hz
Weight465 kg
Machine dimensions (WxDxH)950 x 1200 x 1150 mm

Request your quote

  • By sending this contact form Audion assumes that you are aware of our Privacy Policy

This site uses cookies. Some of these cookies are essential, while others help us to improve your experience by providing insights into how the site is being used.

Privacy Settings saved!
Privacy Settings

When you visit any web site, it may store or retrieve information on your browser, mostly in the form of cookies. Control your personal Cookie Services here.


Analytical cookies help us to improve our website by collecting and reporting information on its usage.
  • _ga
  • _gat
  • _gid

Decline all Services
Accept all Services